On-Site Inspection of Chamber Systems in Ha Nam – Evaluation of Environmental Test Equipment Condition

  • News
  • 94 views

Table of Contents


The technical team of VintechME conducted an on-site inspection of chamber systems in Ha Nam, including a thermal shock chamber, temperature & humidity chamber, and salt spray test chamber. During the inspection, several issues were identified related to the pneumatic system, touchscreen operation, and salt residue build-up caused by long-term inactivity. Periodic inspections help businesses evaluate equipment conditions and ensure more stable system operation.

During the operation of environmental testing equipment, periodic inspections help businesses detect technical issues early, evaluate equipment condition, and minimize impacts on product testing processes.

The technical team of VintechME conducted an on-site inspection of chamber systems in Ha Nam with the following findings:

Thermal Shock Chamber Inspection – High Temperature Chamber Issue

Equipment Condition

The thermal shock chamber was unable to move test samples to the high-temperature chamber during operation.

Inspection Result

After on-site inspection, technicians identified that the issue was caused by air leakage in the pneumatic piping system, resulting in insufficient air pressure and affecting the lifting mechanism that transfers products to the high-temperature chamber.

Inspection Items

  • Inspection of the pneumatic system
  • Inspection of the product lifting mechanism
  • Evaluation of operating air pressure
  • Inspection of high-temperature chamber operation

Some Images of the Chamber Systems:

Temperature & Humidity Chamber Inspection – Touchscreen Failure

Equipment Condition

The temperature and humidity chamber experienced a touchscreen malfunction, preventing operators from controlling and setting operating parameters.

Inspection Result

After checking the control system, technicians determined that the connection cable between the touchscreen and the control board was damaged, causing signal loss.

Inspection Items

  • Inspection of the touchscreen
  • Inspection of connection signals
  • Inspection of signal transmission cables
  • Evaluation of the control board condition

Some Images of the Chamber Systems:

Salt Spray Test Chamber Inspection – Salt Build-Up After Long-Term Inactivity

Equipment Condition

The salt spray test chamber had not been operated for a long period, resulting in salt residue accumulation inside the system, affecting equipment performance and operational stability.

Inspection Result

During the inspection, technicians observed salt deposits accumulated in the piping system and inside the testing chamber due to prolonged inactivity.

The equipment requires periodic maintenance and cleaning to ensure stable operation and minimize impacts on testing performance.

Inspection Items

  • Inspection of the test chamber condition
  • Evaluation of salt residue accumulation
  • Inspection of solution supply lines
  • Inspection of spray system operation

Some Images of the Chamber Systems:

Contact VintechME

If you are experiencing similar issues with your testing chambers, please contact VintechME for fast and accurate technical support:

  • Website: VintechME Website
  • Hotline: 0966 252 565 (Zalo/WhatsApp)
  • Nationwide technical service
  • Experienced engineers specializing in testing equipment
  • Learn more about repair and maintenance services for temperature & humidity chambers provided by VintechME
  • Learn more about environmental test chambers supplied by VintechME

VintechME Vietnam specializes in supplying, repairing, and maintaining testing equipment, environmental temperature & humidity chambers, and large-scale environmental test chambers for QA/QC departments and R&D laboratories.